Detail Process Analyzer For Oxygen By Zirconia Sensor, Model : 413 Series, Brand : Nova
The Nova 413 Series process analyzers utilize a long-life zirconia sensor for high accuracy detection of O2 in process gases. The sensor produces a small millivoltage output which is directly proportional to the Oxygen detected. The output is then linearized and displayed on a digital readout meter and can also be sent to an optional recorder output. The 413 can be supplied with sampling pump or pressure regulator to suit most applications. High and low range versions of this analyzer are available - denoted as models 413 and 413L respectively.
Specifications
Method of Detection
Customer replaceable zirconia O2 sensor. Expected life 5 years or more.
Ranges Available
413 - any range from 0-2.0% to 0-96% O2
413L - any range from 0-200 PPM to 0-20000 PPM O2
Resolution 0.1% O2 or % versions; 1 PPM O2 on PPM versions
Accuracy & Repeatability ± 0.5% of full scale
Drift Less than 0.3% of full scale per month
Response Time ( T-90) 2-3 seconds to 90% step change
Ambient Temperature Range
32 to 122° F ( 0 to 50° C) , Lower temperatures ( -5° F, -20° C) with Cold Weather Package
Linearity ± 0.5% of full scale
Size & Weight Dimensions will vary depending on enclosure style and options required
Power 115 VAC, 60 Hz - 220 VAC, 50 Hz
Output Options 4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
Alarms High and/ or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
Analyzer Type O2 Analyzer
Hazardous Area Rating Options
General Purpose, Non-Hazardous
Applications
For continuous analysis of PPM to 100% oxygen ( O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, oxygen deficiency analysis, and many more.
Features
High accuracy continuous O2 analyzer
Bright digital readout, and 4 - 20 mA outputs
Long life zirconia O2 sensor
Fast response ( T90 - 2 to 3 seconds)
Microprocessor control for push button calibration
Easy-to-maintain modular layout
Built in sample pump or pressure regulator
Options
Hi/ Low gas, low flow, and diagnostic alarms available
Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
Sample conditioning systems available for removal of moisture, acidic gas, and dust
Cabinet purge system available for use in hazardous areas
One-step ' cal-now' button with touch screen LCD display
Full automatic calibration with touch screen LCD display
Cold weather package for operation to -5° F ( -20° C)
Cabinet coolers can be fitted to most models
High temperature filters and probes
Calibration
Nitrogen for zero
Air or analyzed calibration gas for span